On the use of the correction factor with Japanese ozonesonde data
G. A. Morris1, G. Labow2, H. Akimoto3, M. Takigawa4, M. Fujiwara5, F. Hasebe5, J. Hirokawa5, and T. Koide6
1Dept. of Physics and Astronomy, Valparaiso University, Valparaiso, IN, USA
2Science Systems and Applications, Inc., Lanham, MD and NASA Goddard Space Flight Center, Greenbelt, MD, USA
3Asia Center for Air Pollution Research, Niigata, Japan
4Japan Agency for Marine-Earth Science and Technology, Yokohama, Japan
5Faculty of Environmental Earth Sciences, Hokkaido University, Sapporo, Japan
6Atmospheric Environment Division, Japan Meteorological Agency, Tokyo, Japan

Abstract. In submitting data to the World Meteorological Organization (WMO) World Ozone and Ultraviolet Data Center (WOUDC), numerous ozonesonde stations include a correction factor (CF) that multiplies ozone concentration profile data so that the columns computed agree with column measurements from co-located ground-based and/or overpassing satellite instruments. We evaluate this practice through an examination of data from 4 Japanese ozonesonde stations: Kagoshima, Naha, Sapporo, and Tsukuba. While agreement between the sonde columns and Total Ozone Mapping Spectrometer (TOMS) or Ozone Mapping Instrument (OMI) is improved by use of the CF, agreement between the sonde ozone concentrations reported near the surface and data from surface monitors near the launch sites is negatively impacted. In addition, the agreement between the mean sonde columns without the CF and the ground-based Dobson instrument columns is improved by ~1.5% by using the McPeters et al. (1997) balloon burst climatology rather than the constant mixing ratio assumption for the above burst height column estimate. Limited comparisons of coincident ozonesonde profiles from Hokkaido University with those in the WOUDC database suggest that while the application of the CFs in the stratosphere improves agreement, it negatively impacts the agreement in the troposphere. Finally, unexplained trends and changing trends in the CFs appear over the last 20 yr. The overall trend in the CFs for the four Japanese ozonesonde stations from 1990–2010 is (−0.157 ± 0.032) × 10−2 yr−1; but from 1993–1999 the trend is (−2.21 ± 0.14) × 10−2 yr−1 and from 1999–2009 is (1.180 ± 0.059) × 10−2 yr−1, resulting in a statistically significant difference in CF trends between these two periods of (3.39 ± 0.15) × 10−2 yr−1. Given our analysis, we recommend the following: (1) use of the balloon burst climatology is preferred to a constant mixing ratio assumption for determining total column ozone with sonde data; (2) if CFs are applied, their application should probably be restricted to altitudes above the tropopause; (3) only sondes that reach at least 32 km (10.5 hPa) before bursting should be used in data validation and/or ozone trend studies; and (4) all ozone trend studies employing Japanese sonde data should be revisited after a careful examination of the impact of the CF on the calculated ozone trends.

Citation: Morris, G. A., Labow, G., Akimoto, H., Takigawa, M., Fujiwara, M., Hasebe, F., Hirokawa, J., and Koide, T.: On the use of the correction factor with Japanese ozonesonde data, Atmos. Chem. Phys. Discuss., 12, 15597-15638, doi:10.5194/acpd-12-15597-2012, 2012.
 